{"_buckets": {"deposit": "4031f7a4-e456-40b1-9d54-cb8783104883"}, "_deposit": {"created_by": 12, "id": "570", "owners": [12], "pid": {"revision_id": 0, "type": "depid", "value": "570"}, "status": "published"}, "_oai": {"id": "oai:junshin.repo.nii.ac.jp:00000570", "sets": ["93"]}, "author_link": ["1013", "1012", "1008", "1014", "1010", "1011", "1009", "1007"], "item_10002_biblio_info_7": {"attribute_name": "書誌情報", "attribute_value_mlt": [{"bibliographicIssueDates": {"bibliographicIssueDate": "2022-03-31", "bibliographicIssueDateType": "Issued"}, "bibliographicIssueNumber": "12", "bibliographicPageEnd": "075", "bibliographicPageStart": "067", "bibliographic_titles": [{"bibliographic_title": "純真学園大学雑誌"}, {"bibliographic_title": "Journal of Junshin Gakuen University,Faculty of Health Sciences", "bibliographic_titleLang": "en"}]}]}, "item_10002_description_19": {"attribute_name": "フォーマット", "attribute_value_mlt": [{"subitem_description": "application/pdf", "subitem_description_type": "Other"}]}, "item_10002_description_5": {"attribute_name": "抄録", "attribute_value_mlt": [{"subitem_description": "The purpose of this study was to investigate the effect and address the suitable electron energy in cross-calibration with or without an external monitor chamber. First, using the two linear accelerators, the calibration factor (N_(〖D,w,Q〗_cross)^field) and absorbed dose to water calibration factor (N_(D,w)) values are compared with or without an external monitor chamber using cross-calibration in standard dosimetry 12. Next, the N_(〖D,w,Q〗_corss)^field uses five different energies (12, 15, 16, 18, and 20 MeV) for its cross-calibration. Then, 6 MeV electron absorbed dose at depth of the dose maximum (D (dmax)) is calculated. We compared to D (dmax) for the calibration factor (N_(D,w) (Co)) given by 60Co gamma radiation. The N_(〖D,w,Q〗_corss)^field is almost equal in both using and not using an external monitor chamber, the difference value of N_(〖D,w,Q〗_corss)^field was only 0.01%. Between both D (dmax) using calibration factors N_(〖D,w,Q〗_corss)^field and N_(D,w) (Co) is a significant difference (P \u003c 0.001). Moreover, there is also a significant difference between D (dmax) using calibration factor obtained from 20 MeV and each 12 MeV, 15 MeV (P \u003c 0.05). These results demonstrate that there is no significant change with or without the external monitor chamber. Furthermore, the electron energy most suitable for obtaining calibration factors N_(〖D,w,Q〗_corss)^field in cross-calibration is found to be 20 MeV, and the absorbed dose is almost the same value at 12, 15, 16, and 18 MeV.", "subitem_description_type": "Abstract"}]}, "item_10002_publisher_8": {"attribute_name": "出版者", "attribute_value_mlt": [{"subitem_publisher": "純真学園大学"}]}, "item_10002_rights_15": {"attribute_name": "権利", "attribute_value_mlt": [{"subitem_rights": "純真学園大学"}]}, "item_10002_source_id_11": {"attribute_name": "書誌レコードID", "attribute_value_mlt": [{"subitem_source_identifier": "AA12581584", "subitem_source_identifier_type": "NCID"}]}, "item_10002_source_id_9": {"attribute_name": "ISSN", "attribute_value_mlt": [{"subitem_source_identifier": "2186-6481", "subitem_source_identifier_type": "ISSN"}]}, "item_10002_version_type_20": {"attribute_name": "著者版フラグ", "attribute_value_mlt": [{"subitem_version_resource": "http://purl.org/coar/version/c_970fb48d4fbd8a85", "subitem_version_type": "VoR"}]}, "item_creator": {"attribute_name": "著者", "attribute_type": "creator", "attribute_value_mlt": [{"creatorNames": [{"creatorName": "久家, 教幸"}, {"creatorName": "クガ, ノリユキ", "creatorNameLang": "ja-Kana"}], "nameIdentifiers": [{"nameIdentifier": "1007", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "尻枝, 勝敏"}, {"creatorName": "シリエダ, カツトシ", "creatorNameLang": "ja-Kana"}], "nameIdentifiers": [{"nameIdentifier": "1008", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "下田平, 春彦"}, {"creatorName": "シモタビラ, ハルヒコ", "creatorNameLang": "ja-Kana"}], "nameIdentifiers": [{"nameIdentifier": "1009", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "下野, 哲範"}, {"creatorName": "シモノ, テツノリ", "creatorNameLang": "ja-Kana"}], "nameIdentifiers": [{"nameIdentifier": "1010", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "KUGA, Noriyuki ", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "1011", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "SHIRIEDA, Katsutoshi ", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "1012", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "SHIMOTABIRA, Haruhiko", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "1013", "nameIdentifierScheme": "WEKO"}]}, {"creatorNames": [{"creatorName": "SHIMONO, Tetsunori ", "creatorNameLang": "en"}], "nameIdentifiers": [{"nameIdentifier": "1014", "nameIdentifierScheme": "WEKO"}]}]}, "item_files": {"attribute_name": "ファイル情報", "attribute_type": "file", "attribute_value_mlt": [{"accessrole": "open_date", "date": [{"dateType": "Available", "dateValue": "2022-03-31"}], "displaytype": "detail", "download_preview_message": "", "file_order": 0, "filename": "純真学園大学雑誌第12号_P059-065.pdf", "filesize": [{"value": "1.8 MB"}], "format": "application/pdf", "future_date_message": "", "is_thumbnail": false, "licensetype": "license_11", "mimetype": "application/pdf", "size": 1800000.0, "url": {"label": "Effect of Electron Beam Energies on the Calibration Factor in Cross-Calibration With and Without an External Monitor Chamber", "url": "https://junshin.repo.nii.ac.jp/record/570/files/純真学園大学雑誌第12号_P059-065.pdf"}, "version_id": "eb38248e-00ae-4fdf-bdfd-2ffe3bc457e5"}]}, "item_keyword": {"attribute_name": "キーワード", "attribute_value_mlt": [{"subitem_subject": "cross-calibration", "subitem_subject_scheme": "Other"}, {"subitem_subject": "electron beam dosimetry", "subitem_subject_scheme": "Other"}, {"subitem_subject": "electron beam energy", "subitem_subject_scheme": "Other"}, {"subitem_subject": "external monitor chamber", "subitem_subject_scheme": "Other"}, {"subitem_subject": "cross-calibration", "subitem_subject_language": "en", "subitem_subject_scheme": "Other"}, {"subitem_subject": "electron beam dosimetry", "subitem_subject_language": "en", "subitem_subject_scheme": "Other"}, {"subitem_subject": "electron beam energy", "subitem_subject_language": "en", "subitem_subject_scheme": "Other"}, {"subitem_subject": "external monitor chamber", "subitem_subject_language": "en", "subitem_subject_scheme": "Other"}]}, "item_language": {"attribute_name": "言語", "attribute_value_mlt": [{"subitem_language": "eng"}]}, "item_resource_type": {"attribute_name": "資源タイプ", "attribute_value_mlt": [{"resourcetype": "departmental bulletin paper", "resourceuri": "http://purl.org/coar/resource_type/c_6501"}]}, "item_title": "Effect of Electron Beam Energies on the Calibration Factor in Cross-Calibration With and Without an External Monitor Chamber", "item_titles": {"attribute_name": "タイトル", "attribute_value_mlt": [{"subitem_title": "Effect of Electron Beam Energies on the Calibration Factor in Cross-Calibration With and Without an External Monitor Chamber"}, {"subitem_title": "Effect of Electron Beam Energies on the Calibration Factor in Cross-Calibration With and Without an External Monitor Chamber", "subitem_title_language": "en"}]}, "item_type_id": "10002", "owner": "12", "path": ["93"], "permalink_uri": "https://junshin.repo.nii.ac.jp/records/570", "pubdate": {"attribute_name": "公開日", "attribute_value": "2022-03-31"}, "publish_date": "2022-03-31", "publish_status": "0", "recid": "570", "relation": {}, "relation_version_is_last": true, "title": ["Effect of Electron Beam Energies on the Calibration Factor in Cross-Calibration With and Without an External Monitor Chamber"], "weko_shared_id": -1}
Effect of Electron Beam Energies on the Calibration Factor in Cross-Calibration With and Without an External Monitor Chamber
https://junshin.repo.nii.ac.jp/records/570
https://junshin.repo.nii.ac.jp/records/570979c1aa0-f920-4be2-87f0-d8df9ede61b7
名前 / ファイル | ライセンス | アクション |
---|---|---|
Effect of Electron Beam Energies on the Calibration Factor in Cross-Calibration With and Without an External Monitor Chamber (1.8 MB)
|
Item type | 紀要論文 / Departmental Bulletin Paper(1) | |||||
---|---|---|---|---|---|---|
公開日 | 2022-03-31 | |||||
言語 | ||||||
言語 | eng | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | cross-calibration | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | electron beam dosimetry | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | electron beam energy | |||||
キーワード | ||||||
主題Scheme | Other | |||||
主題 | external monitor chamber | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | cross-calibration | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | electron beam dosimetry | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | electron beam energy | |||||
キーワード | ||||||
言語 | en | |||||
主題Scheme | Other | |||||
主題 | external monitor chamber | |||||
資源タイプ | ||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||
資源タイプ | departmental bulletin paper | |||||
著者 |
久家, 教幸
× 久家, 教幸× 尻枝, 勝敏× 下田平, 春彦× 下野, 哲範× KUGA, Noriyuki× SHIRIEDA, Katsutoshi× SHIMOTABIRA, Haruhiko× SHIMONO, Tetsunori |
|||||
抄録 | ||||||
内容記述タイプ | Abstract | |||||
内容記述 | The purpose of this study was to investigate the effect and address the suitable electron energy in cross-calibration with or without an external monitor chamber. First, using the two linear accelerators, the calibration factor (N_(〖D,w,Q〗_cross)^field) and absorbed dose to water calibration factor (N_(D,w)) values are compared with or without an external monitor chamber using cross-calibration in standard dosimetry 12. Next, the N_(〖D,w,Q〗_corss)^field uses five different energies (12, 15, 16, 18, and 20 MeV) for its cross-calibration. Then, 6 MeV electron absorbed dose at depth of the dose maximum (D (dmax)) is calculated. We compared to D (dmax) for the calibration factor (N_(D,w) (Co)) given by 60Co gamma radiation. The N_(〖D,w,Q〗_corss)^field is almost equal in both using and not using an external monitor chamber, the difference value of N_(〖D,w,Q〗_corss)^field was only 0.01%. Between both D (dmax) using calibration factors N_(〖D,w,Q〗_corss)^field and N_(D,w) (Co) is a significant difference (P < 0.001). Moreover, there is also a significant difference between D (dmax) using calibration factor obtained from 20 MeV and each 12 MeV, 15 MeV (P < 0.05). These results demonstrate that there is no significant change with or without the external monitor chamber. Furthermore, the electron energy most suitable for obtaining calibration factors N_(〖D,w,Q〗_corss)^field in cross-calibration is found to be 20 MeV, and the absorbed dose is almost the same value at 12, 15, 16, and 18 MeV. | |||||
書誌情報 |
純真学園大学雑誌 en : Journal of Junshin Gakuen University,Faculty of Health Sciences 号 12, p. 067-075, 発行日 2022-03-31 |
|||||
出版者 | ||||||
出版者 | 純真学園大学 | |||||
ISSN | ||||||
収録物識別子タイプ | ISSN | |||||
収録物識別子 | 2186-6481 | |||||
書誌レコードID | ||||||
収録物識別子タイプ | NCID | |||||
収録物識別子 | AA12581584 | |||||
権利 | ||||||
権利情報 | 純真学園大学 | |||||
フォーマット | ||||||
内容記述タイプ | Other | |||||
内容記述 | application/pdf | |||||
著者版フラグ | ||||||
出版タイプ | VoR | |||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 |