【原著】Effect of Electron Beam Energies on the Calibration Factor in Cross-Calibration With and Without an External Monitor Chamber
https://junshin.repo.nii.ac.jp/records/570
https://junshin.repo.nii.ac.jp/records/570979c1aa0-f920-4be2-87f0-d8df9ede61b7
名前 / ファイル | ライセンス | アクション |
---|---|---|
![]() |
Item type | 紀要論文 / Departmental Bulletin Paper(1) | |||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
公開日 | 2022-03-31 | |||||||||||||||||||||||||||||
タイトル | ||||||||||||||||||||||||||||||
タイトル | 【原著】Effect of Electron Beam Energies on the Calibration Factor in Cross-Calibration With and Without an External Monitor Chamber | |||||||||||||||||||||||||||||
言語 | en | |||||||||||||||||||||||||||||
言語 | ||||||||||||||||||||||||||||||
言語 | eng | |||||||||||||||||||||||||||||
キーワード | ||||||||||||||||||||||||||||||
主題Scheme | Other | |||||||||||||||||||||||||||||
主題 | cross-calibration | |||||||||||||||||||||||||||||
キーワード | ||||||||||||||||||||||||||||||
主題Scheme | Other | |||||||||||||||||||||||||||||
主題 | electron beam dosimetry | |||||||||||||||||||||||||||||
キーワード | ||||||||||||||||||||||||||||||
主題Scheme | Other | |||||||||||||||||||||||||||||
主題 | electron beam energy | |||||||||||||||||||||||||||||
キーワード | ||||||||||||||||||||||||||||||
主題Scheme | Other | |||||||||||||||||||||||||||||
主題 | external monitor chamber | |||||||||||||||||||||||||||||
キーワード | ||||||||||||||||||||||||||||||
言語 | en | |||||||||||||||||||||||||||||
主題Scheme | Other | |||||||||||||||||||||||||||||
主題 | cross-calibration | |||||||||||||||||||||||||||||
キーワード | ||||||||||||||||||||||||||||||
言語 | en | |||||||||||||||||||||||||||||
主題Scheme | Other | |||||||||||||||||||||||||||||
主題 | electron beam dosimetry | |||||||||||||||||||||||||||||
キーワード | ||||||||||||||||||||||||||||||
言語 | en | |||||||||||||||||||||||||||||
主題Scheme | Other | |||||||||||||||||||||||||||||
主題 | electron beam energy | |||||||||||||||||||||||||||||
キーワード | ||||||||||||||||||||||||||||||
言語 | en | |||||||||||||||||||||||||||||
主題Scheme | Other | |||||||||||||||||||||||||||||
主題 | external monitor chamber | |||||||||||||||||||||||||||||
資源タイプ | ||||||||||||||||||||||||||||||
資源タイプ識別子 | http://purl.org/coar/resource_type/c_6501 | |||||||||||||||||||||||||||||
資源タイプ | departmental bulletin paper | |||||||||||||||||||||||||||||
著者 |
久家, 教幸
× 久家, 教幸
× 尻枝, 勝敏
× 下田平, 春彦
× 下野, 哲範
× KUGA, Noriyuki
× SHIRIEDA, Katsutoshi
× SHIMOTABIRA, Haruhiko
× SHIMONO, Tetsunori
|
|||||||||||||||||||||||||||||
抄録 | ||||||||||||||||||||||||||||||
内容記述タイプ | Abstract | |||||||||||||||||||||||||||||
内容記述 | The purpose of this study was to investigate the effect and address the suitable electron energy in cross-calibration with or without an external monitor chamber. First, using the two linear accelerators, the calibration factor (N_(〖D,w,Q〗_cross)^field) and absorbed dose to water calibration factor (N_(D,w)) values are compared with or without an external monitor chamber using cross-calibration in standard dosimetry 12. Next, the N_(〖D,w,Q〗_corss)^field uses five different energies (12, 15, 16, 18, and 20 MeV) for its cross-calibration. Then, 6 MeV electron absorbed dose at depth of the dose maximum (D (dmax)) is calculated. We compared to D (dmax) for the calibration factor (N_(D,w) (Co)) given by 60Co gamma radiation. The N_(〖D,w,Q〗_corss)^field is almost equal in both using and not using an external monitor chamber, the difference value of N_(〖D,w,Q〗_corss)^field was only 0.01%. Between both D (dmax) using calibration factors N_(〖D,w,Q〗_corss)^field and N_(D,w) (Co) is a significant difference (P < 0.001). Moreover, there is also a significant difference between D (dmax) using calibration factor obtained from 20 MeV and each 12 MeV, 15 MeV (P < 0.05). These results demonstrate that there is no significant change with or without the external monitor chamber. Furthermore, the electron energy most suitable for obtaining calibration factors N_(〖D,w,Q〗_corss)^field in cross-calibration is found to be 20 MeV, and the absorbed dose is almost the same value at 12, 15, 16, and 18 MeV. | |||||||||||||||||||||||||||||
書誌情報 |
純真学園大学雑誌 en : Journal of Junshin Gakuen University,Faculty of Health Sciences 号 12, p. 067-075, 発行日 2022-03-31 |
|||||||||||||||||||||||||||||
出版者 | ||||||||||||||||||||||||||||||
出版者 | 純真学園大学 | |||||||||||||||||||||||||||||
ISSN | ||||||||||||||||||||||||||||||
収録物識別子タイプ | ISSN | |||||||||||||||||||||||||||||
収録物識別子 | 2186-6481 | |||||||||||||||||||||||||||||
書誌レコードID | ||||||||||||||||||||||||||||||
収録物識別子タイプ | NCID | |||||||||||||||||||||||||||||
収録物識別子 | AA12581584 | |||||||||||||||||||||||||||||
権利 | ||||||||||||||||||||||||||||||
権利情報 | 純真学園大学 | |||||||||||||||||||||||||||||
フォーマット | ||||||||||||||||||||||||||||||
内容記述タイプ | Other | |||||||||||||||||||||||||||||
内容記述 | application/pdf | |||||||||||||||||||||||||||||
著者版フラグ | ||||||||||||||||||||||||||||||
出版タイプ | VoR | |||||||||||||||||||||||||||||
出版タイプResource | http://purl.org/coar/version/c_970fb48d4fbd8a85 |